IATF 16949 asks for the same things as ISO 9001 – more precisely, and with a few extras: as-found records, laboratory scope, and a measurement system study for every gauge in the control plan. Here is our reading, what auditors look for, and the findings we see most often.

Last updated: November 6, 2026 · Reviewed by irot

  • Free for up to 20 devices
  • No payment details
  • Interface in English and German
MSA
MSA study in Kalibrat

What IATF 16949 adds to ISO 9001 for measuring equipment

If you already meet the measuring-equipment requirements of ISO 9001, IATF 16949 asks for the same thing – only more precisely, and with a few extras. In our reading, the additions for measuring equipment come down to six points:

Records that show what happened
Calibration and verification records must include revisions after engineering changes, the as-found condition when a device comes in for calibration, and a statement of conformity afterwards – not only a list of readings.
Out-of-spec handling
If a device was found outside its specification, you assess the risk to product already measured with it. If suspect product has been shipped, the customer is told.
Internal laboratory
If you calibrate in-house, you define and document the scope: which methods, which devices, who is competent, which reference standards you use and how they are traced.
External laboratory
Outside calibration normally comes from a laboratory with accredited scope for that measurement (typically ISO/IEC 17025) – or from one that your customer has approved in writing. The certificate should make the scope visible.
Measurement system analysis
For every measuring system named in the control plan, a statistical study of the variation in its results. IATF 16949 clause 7.1.5.1.1 is the usual reference.
Customer-specific requirements
Individual customers add their own acceptance limits and formats. They override our general summary.

This is a summary in our own words, not the wording of the standard. IATF 16949, the AIAG MSA manual and your customer's requirements are binding.

What auditors look for

  1. The link to the control plan. Pick a characteristic, find the gauge that measures it, and follow the trail to its calibration record and its measurement system study.
  2. The as-found condition. Not only “calibrated, OK”, but what the device showed when it arrived.
  3. The assessment. One real example of an out-of-tolerance finding, with the parts, batches or orders you looked at and the conclusion.
  4. The laboratory. Whose scope covered this measurement, and where is that shown on the certificate?
  5. History. The same device, three calibrations back, still readable.

The records you need to be able to produce

  • Device ID, location or owner, and the characteristic it is used for.
  • For every calibration: date, result, who calibrated, the certificate and its number, and the laboratory's accreditation reference where relevant. See what a calibration certificate must show.
  • The as-found result and the decision after it.
  • For each out-of-tolerance result: the assessment of earlier measurements and who signed it off.
  • Records of repairs and the verification before the device returned to use.
  • The measurement system study for each gauge named in the control plan.
  • The interval and the reasoning behind it – see calibration intervals.

Measurement system analysis in one paragraph

Calibration asks whether the device reads correctly. A measurement system study asks something different: is the whole measuring process – device, operator, fixture, method, environment – good enough for this tolerance? A calibrated calliper can still be useless for a ±0.01 mm feature. The two common studies are a capability study on a reference standard (Cg and Cgk – try the Cgk calculator) and a gauge repeatability and reproducibility study with several operators and parts, which gives %GRR and the number of distinct categories (try the gauge R&R calculator). Which method and which limits your customer accepts is the customer's call – check their requirements before you plan the study.

Common findings

  • Gauges in the control plan without any measurement system study – or a study on a different gauge or a different tolerance.
  • Certificates that show the condition after adjustment only; the as-found values were never requested.
  • An external laboratory whose accreditation does not cover the quantity or range that was calibrated.
  • An out-of-tolerance result with no written assessment of earlier measurements.
  • A study that was “done once” years ago and never repeated after a change of fixture, operator group or tolerance.
  • Spreadsheets where a corrected entry replaced the original, so nobody can tell what was recorded at the time.

IATF 16949 records in Kalibrat

Kalibrat is software that holds your records. It does not make you compliant with IATF 16949 – that remains your organisation's job – and it is not accredited or certified for anything. What the software does:

  • Each calibration record carries the certificate number and the laboratory's accreditation reference, plus fields for the deviation found and the assessment of earlier results.
  • For a measuring device, an out-of-tolerance result cannot be saved without that assessment, and a pre-filled draft must be replaced with your own assessment.
  • With the IATF 16949 add-on module, measurement system studies (capability study, gauge R&R with and without operators) are stored per device, each with its characteristic and tolerance, and show %GRR and ndc. A completed study is frozen as calculated.
  • Records are not overwritten: a correction cancels the entry with a reason, and the original stays visible.
  • Every change goes into a change log in which each entry is chained to the one before it.

Which devices need a study, and whether the result is acceptable to your customer, stays with you. The device register itself is free for up to 20 devices – try it with your own gauges.

FAQ

Frequently asked questions

Does IATF 16949 require an accredited calibration laboratory?

In our reading, external calibration normally comes from a laboratory with accredited scope for that measurement, or from one your customer has approved. In-house calibration needs a documented laboratory scope.

Which gauges need a measurement system study?

Those named in the control plan – the measurement systems that decide whether a part is good or bad. Your customer may extend the list.

Is a gauge R&R study enough?

It answers one question: how much of the tolerance the measurement process uses. A capability study on a reference (Cg, Cgk) answers another. Which studies and limits apply is the customer's call.

Hold the records the audit asks for – the register is free up to 20 devices