Calibration certificate · as found / as left How to read a calibration certificate

Most certificates are filed unread. Here is which lines matter, what “as found” and “as left” tell you, and whose decision pass or fail really is – and what to do when a gauge comes back out of tolerance.

Last updated: November 6, 2026 · Reviewed by irot

  • Free for up to 20 devices
  • No payment details
  • Interface in English and German
Device
Calibration record with certificate in Kalibrat

What a calibration certificate is for

A calibration certificate is the written proof that a measuring instrument was compared with a better one, and by how much it differed. It is not a pass mark: it says what the laboratory found, and whether that is good enough for your measurement is still your decision. Most certificates are filed unread – which is why they fail in the audit.

The lines that matter

Identification
Certificate number, laboratory, and the instrument itself: manufacturer, type, serial number, often your own ID as well. Compare the serial number with the gauge in your hand before anything else. Certificates get swapped when twenty gauges come back in one box, and a certificate for the wrong serial number proves nothing for this gauge.
Reference standards and traceability
Which standards the laboratory used, and a statement that they link back to national or international standards. Accredited laboratories usually print their accreditation number as well; you can look it up with the accreditation body, together with the scope it covers.
Results
For each measuring point: the nominal value, the reading and the deviation. If the laboratory adjusted the instrument, a careful certificate shows the state as found (before adjustment) and as left (after). The as-found figures are the ones that tell you whether past measurements were safe.
Measurement uncertainty
Usually the expanded uncertainty with a coverage factor of 2, which covers roughly 95 % of the values the true result could take. The smaller it is compared with your tolerance, the more the figures can be trusted.
Environmental conditions
Temperature, sometimes humidity. Length measurements refer to 20 °C; a certificate that does not say where the laboratory measured leaves a question open.
Dates and signature
The calibration date is not the issue date. A “next calibration recommended” line is the laboratory's suggestion – the interval is your decision and you should be able to justify it. Look for a named, authorised signatory.

Certificate, report or sticker?

The words are used loosely, but the three things differ.

  • Certificate – the formal document from an external laboratory, usually accredited, with traceability and uncertainty. This is what auditors ask for when the instrument is a reference or a critical gauge.
  • Report – the record of a calibration, often done in-house against your own reference standard. It is valid if the standard is itself traceable and the method is written down.
  • Sticker – only a pointer to the record, saying when the next calibration is due. It is not evidence by itself. A sticker without a certificate or report behind it is a claim, not a proof.

Pass or fail: whose decision?

A certificate may carry a statement of conformity – “within specification” – and then it should also name the decision rule behind it. Two common approaches give different answers for a reading close to the limit. With simple acceptance the reading alone decides, which is only reasonable when the uncertainty is small compared with the tolerance (a ratio of 4 : 1 is the usual rule of thumb). With a guard band, the reading plus its uncertainty must stay inside the limit; in between lies a grey zone in which compliance can be neither proved nor disproved. If the certificate gives no statement, apply your own limit and your own rule – and write down which one.

When the result is out of tolerance

  • Block the instrument until you know what happened. Do not wait for the next scheduled use.
  • Look at as-found, not as-left. An adjustment that brings the gauge back into tolerance does not undo what it measured while it was off.
  • Assess earlier measurements. ISO 9001 expects you to judge whether results obtained with the instrument are still valid: since when has it been drifting, which parts were measured against which tolerance, who had it. That judgement needs someone who knows the parts; software can only collect the facts.
  • Act on the finding – recall, rework, inform the customer where required – and consider a shorter interval.

File it where it can be found

In an audit nobody asks for “certificate 4711”. They ask for the certificate of the micrometer lying on the table. File it with the instrument, not in a folder sorted by date or in a mailbox, and make sure certificate number and serial number are readable in the record.

Certificates and results in Kalibrat

In Kalibrat each calibration record has fields for the certificate number and the laboratory's accreditation number, and the certificate itself can be attached to that record as a file. Measured values are stored per point with nominal value, deviation, limit and – where you have it – the expanded uncertainty (k = 2). The record also keeps the reference standard used, an optional room temperature and the decision rule applied, so an old result is not re-judged when you later change the setting. You choose between a guard-band rule and simple acceptance; points whose uncertainty makes the result undecidable are shown as such. An “out of tolerance” result cannot be saved without your assessment of earlier measurements. An audit check lists external calibrations that have no certificate number, no accreditation entry or no attached document. Free for up to 20 devices – try it with your own gauges.

FAQ

Frequently asked questions

What is the difference between a calibration certificate and a calibration report?

A certificate is the formal document from an external, usually accredited, laboratory with traceability and uncertainty. A report records a calibration, often done in-house against your own traceable reference. Both are evidence; the sticker on the gauge is not.

What does “as found” mean on a calibration certificate?

The state of the instrument before any adjustment. It tells you whether the measurements made before the calibration were safe. “As left” is the state after adjustment.

Does a calibration certificate say pass or fail?

Only if it carries a statement of conformity, and then it should name the decision rule. Otherwise you compare the deviation with your own limit, allowing for the uncertainty.

How long do I have to keep calibration certificates?

As long as the measurements made with the instrument may be questioned – in practice for the life of the product records they support. Your quality system or customer contracts set the period.