In-house calibration · ISO/IEC 17025 ISO/IEC 17025 and in-house calibration: the records that hold up.
Most in-house calibrations do not need an accredited lab – but they need records an auditor can follow: which reference, which conditions, which uncertainty, which decision rule. Here is what belongs in them.
Last updated: November 13, 2026 · Reviewed by irot
- Free for up to 20 devices
- No payment details
- Interface in English and German

ISO/IEC 17025 is the standard for the competence of testing and calibration laboratories. Accreditation bodies assess labs against it, and an accredited calibration certificate is what most quality managers want for their reference standards. Many manufacturers also calibrate their everyday gauges in-house. This page explains, in our own words, how the two fit together and which records an in-house calibration should keep if you want them to hold up – whether or not you ever seek accreditation.
Kalibrat is software for managing calibration records. It is not a laboratory, it is not accredited, and using it does not make your lab accredited or your calibrations conform to ISO/IEC 17025. That judgement belongs to your accreditation body or your auditor.
Do in-house calibrations need ISO/IEC 17025?
For most manufacturers, no. ISO 9001 asks for measuring equipment that is calibrated or verified against standards traceable to national or international standards; it does not ask for an accredited lab. A machine shop that checks its calipers against its own gauge blocks meets that, provided the gauge blocks themselves carry a valid traceable certificate and the in-house procedure is documented.
Automotive is stricter. In our reading, IATF 16949 expects an in-house laboratory to have a defined scope and to show it is technically competent for what it calibrates – and external calibrations should, as a rule, come from labs accredited to ISO/IEC 17025 for that calibration. Details are on our page about IATF 16949 calibration requirements. Your customer's requirements may go further; they are binding, not this page.
What a solid in-house calibration record contains
Whether or not you work towards accreditation, the habits of a 17025 lab are a good model for in-house records. A record that an auditor can follow typically answers these questions:
- Which item? Unique ID and serial number, so the record cannot be confused with a similar gauge.
- When, and by whom? Date of calibration and the person who did it – and that person's qualification for the task.
- How? The procedure or work instruction used, so the next calibration is comparable.
- Against what? The reference standard used, with its own ID and certificate – this is where traceability is shown.
- Under which conditions? For dimensional work especially the room temperature, because a few degrees change the result measurably.
- What was measured? Nominal value, reading and deviation at each calibration point, not only “OK”.
- How certain is it? The measurement uncertainty, usually stated as expanded uncertainty with coverage factor k = 2.
- What was decided, by which rule? Conforming or not, and the decision rule used – whether the uncertainty was taken into account and how.
- What happened next? Adjustment, repair, a shorter interval – and if the gauge was out of tolerance, the assessment of earlier measurements.
Measurement uncertainty and decision rules
The step most in-house calibrations skip is the uncertainty. Without it, a reading just inside the limit looks like a pass when the true value may well be outside. A decision rule says what you do about that: accept only when the reading plus its uncertainty is inside the limit (guard band), or accept on the reading alone and state that you did. Either can be legitimate; what matters is that the rule is defined in advance and recorded with each result. Our free measurement uncertainty calculator builds a simple budget, and the decision rule calculator shows how the verdict changes.
How Kalibrat keeps these records
- Each calibration record holds the date, the person, the result, and the certificate as an attached file; external calibrations have fields for the certificate number and the lab's accreditation number.
- The record keeps the reference standard used and, optionally, the room temperature at the time of measurement.
- With the Metrology (€5 per month) module, measured values are stored per point with the expanded uncertainty (k = 2) and judged by the decision rule you choose – guard band or simple acceptance – and the rule is stored with the result, so an old result is not re-judged when the setting changes later. The same module shows the traceability chain in both directions and drift per point.
- An “out of tolerance” result cannot be saved without an assessment of earlier measurements; Kalibrat drafts it from the data, you complete it.
- Recorded calibrations cannot be overwritten. A correction cancels the entry with a reason and records anew; every change is in a chained log.
- With the Audit & evidence (€8 per month) module, the audit check lists calibrations by people whose qualification had expired, external calibrations without certificate number or accreditation entry, and traceability chains with a gap. With the Lab & calibration plan (€5 per month) module you can rate the external labs you use.
What Kalibrat does not do here
- It does not issue accredited calibration certificates and does not replace an accredited lab.
- It does not validate your calibration procedures. The uncertainty calculator in the Metrology module helps with a budget, but which contributions belong in it is your lab's decision.
- It does not decide whether your in-house lab is competent; that is assessed by your auditor or accreditation body.
Frequently asked questions
Do I need an ISO/IEC 17025 accredited lab for in-house calibration?
Usually not. ISO 9001 asks for traceable calibration, not accreditation. Your reference standards should carry valid traceable certificates, and the in-house procedure should be documented. Automotive customers and IATF 16949 can ask for more.
Is Kalibrat ISO/IEC 17025 certified or accredited?
No. Kalibrat is software for calibration records. It is not a laboratory and holds no accreditation; using it does not make a lab accredited.
What is a decision rule in calibration?
The rule that says how measurement uncertainty is taken into account when you declare a result conforming – for example accepting only when the reading plus its uncertainty is inside the limit.
Which environmental conditions should an in-house calibration record?
For dimensional calibration at least the room temperature, because length changes with temperature. Other methods may need humidity or other conditions; your procedure defines them.
Keep every calibration record complete – free up to 20 devices