Calibration intervals How often should measuring equipment be calibrated?

The honest answer is: as often as your results say – starting from a sensible default. Here are typical starting intervals for common gauges and a simple way to adjust them from your own calibration results.

Last updated: November 6, 2026 · Reviewed by irot

  • Free for up to 20 devices
  • No payment details
  • Interface in English and German
Overview
Due dates in Kalibrat

There is no single right interval

No standard tells you that a caliper must be calibrated every twelve months. ISO 9001 and IATF 16949 ask for calibration at specified intervals – you specify them, and you have to be able to explain why. A good interval is short enough that a drifting gauge is caught before it has measured many parts, and long enough that you do not pay for calibrations that never find anything.

Typical starting intervals

Where you have no history yet, these are common starting points. They are the values Kalibrat suggests when you create a device – guide values, not requirements. Manufacturer instructions, customer requirements and your own results take precedence.

Measuring equipmentTypical start
Caliper1 year
External micrometer1 year
Internal micrometer (bore gauge)1 year
Dial indicator1 year
Fine indicator / lever dial indicator1 year
Depth caliper / depth gauge1 year
Height gauge / measuring stand1 year
Gauge block / gauge block set3 years
Plug gauge / pin gauge1 year
Ring gauge / setting ring2 years
Thread plug gauge / thread ring gauge1 year
Knife-edge square / try square / protractor2 years
Surface plate / granite surface plate3 years
Precision spirit level / inclinometer2 years
Feeler gauge / radius gauge2 years
Measuring probe / 3D probe / special measuring equipment1 year
Coordinate measuring machine (CMM)1 year
Surface roughness tester1 year
Hardness tester1 year
Torque wrench / torque meter1 year
Scales2 years
Thermometer / climate sensor for the metrology room2 years
Pressure gauge / manometer1 year
Multimeter / electrical test instrument2 years

Guide values from Kalibrat's device catalogue (the same values the application pre-fills). They are not taken from a regulation; where a regulation or your customer sets an interval, that interval applies.

What should change the interval

  • How often and how hard the gauge is used. A caliper used all day at the machine wears faster than one in the inspection room.
  • Environment. Coolant, dust, temperature swings and transport shorten intervals.
  • How critical the measurement is. A gauge that releases safety-relevant parts deserves a shorter interval than one used for rough checks.
  • The manufacturer's recommendation – especially for electronic and force-measuring instruments.
  • Your as-found results. This is the strongest evidence you have.

Adjusting intervals from your own results

Every calibration certificate tells you the as-found condition: was the gauge within tolerance when it arrived at the lab? Use it.

  1. Found out of tolerance: shorten the interval – and check what was measured with the gauge since its last good calibration. That check is the part auditors look for.
  2. Found in tolerance, but close to the limit: keep the interval and watch the trend.
  3. Found well within tolerance several times in a row: you may extend the interval in steps. Document the decision and the results it rests on.

Usage-based intervals work too: a torque wrench, for example, can be due after a number of cycles or after a period, whichever comes first.

Intermediate checks

Between two calibrations a short check against a reference – a gauge block, a setting ring, a check weight – catches damage early. It does not replace calibration, but it can justify a longer interval and tells you on the day, not months later, that something is wrong.

Keeping it under control

The interval is only as good as the reminder behind it. In Kalibrat each device carries its interval in months or weeks, the due date follows from the last calibration, the responsible person is reminded in advance, and each result is recorded with its certificate, so the next interval decision has its evidence attached. Free for up to 20 devices.

FAQ

Frequently asked questions

How often should a caliper be calibrated?

A common starting point is once a year. Shorten it if the caliper is used heavily or found out of tolerance; extend it in steps if several calibrations in a row find it well within tolerance.

Who decides the calibration interval?

Your company does, unless a regulation, the manufacturer or your customer specifies one. ISO 9001 asks for specified intervals and for evidence – not for a particular number.

Can I extend a calibration interval?

Yes, if your records show that the device was repeatedly found well within tolerance. Extend in steps and document the decision with the results it is based on.